Two-dimensional Fourier transform ion cyclotron resonance mass spectrometry : reduction of scintillation noise using Cadzow data processing

Date de publication: 3 mai 2011

M. A. van Agthoven, M.-A. Coutouly, C. Rolando, M.-A. Delsuc
Rapid Commun. Mass Spectrom. 25 1609–1616 (2011). DOI

Travail réalisé sur le site de l’Université de Lille 1, Sciences et Technologies.

Abstract

In two-dimensional Fourier transform ion cyclotron resonance mass spectrometry (2D FTICR-MS), scintillation noise, caused mostly by fluctuations in the number of ions in the ICR cell, is the leading cause for errors in spectrum interpretation. In this study, we adapted an algorithm based on singular value decomposition and first introduced by Cadzow et al. (IEE Proceedings Pt. F 1987, 134, 69) to 2D FTICR-MS and we measured its performance in terms of noise reduction without losing signal information in the 2D mass spectrum